Difference between revisions of "Edwards APG-L"

From Electronen Mikroskopie Museum Nürnberg
Jump to navigation Jump to search
(Basic data of the APG-L gauge head)
 
m (corrected syntax)
Line 1: Line 1:
 
The APG-L is a early RJ45 connected active Pirani Gauge from [[BOC Edwards]].
 
The APG-L is a early RJ45 connected active Pirani Gauge from [[BOC Edwards]].
  
10 to 1 x 10-4 mbar
 
 
{| class="wikitable" style="margin:auto"
 
{| class="wikitable" style="margin:auto"
 
|+ Specifications <ref> [https://www.idealvac.com/files/brochures/Edwards_APG_MP_Manual.pdf | Edwards_APG_MP_Manual.pdf, idealvac.com, published by Edwards Limited, 2010] </ref>
 
|+ Specifications <ref> [https://www.idealvac.com/files/brochures/Edwards_APG_MP_Manual.pdf | Edwards_APG_MP_Manual.pdf, idealvac.com, published by Edwards Limited, 2010] </ref>
 
 
|-
 
|-
| Maximum internal pressure: | 6 bar absolute (5 bar gauge)
+
| Maximum internal pressure: || 6 bar absolute (5 bar gauge)
 
|-
 
|-
| Pressure Range: | 10 to 1 x 10⁻⁴ mbar
+
| Pressure Range: || 10 to 1 x 10⁻⁴ mbar
 
|-
 
|-
| Volume of gauge tube: | 6 cm³  (NW16/NW25) / 10 cm³ (15 mm OD)
+
| Volume of gauge tube: || 6 cm³  (NW16/NW25) / 10 cm³ (15 mm OD)
 
|-
 
|-
| Maximum operating altitude: | 2000 m (Indoor use only)
+
| Maximum operating altitude: || 2000 m (Indoor use only)
 
|-  
 
|-  
|Filament operating temperature (approximate): | 100 ºC
+
|Filament operating temperature (approximate): || 100 ºC
 
|-
 
|-
| Supply Voltage: | +20 to +36V DC (<1V Ripple, <100 Ω Source)
+
| Supply Voltage: || +20 to +36V DC (<1V Ripple, <100 Ω Source)
 
|-
 
|-
| Maximum power consumption: | 1.5W
+
| Maximum power consumption: || 1.5W
 
|-  
 
|-  
| Signal Range: | 2 ≤ output ≤ 10 V DC
+
| Signal Range: || 2 ≤ output ≤ 10 V DC
 
|-
 
|-
| Min. Load: | 10kΩ
+
| Min. Load: || 10kΩ
 
|-
 
|-
| Set-point Adjustment Range: | 2.2 to 8.9 V
+
| Set-point Adjustment Range: || 2.2 to 8.9 V
 
|-
 
|-
| Set-point External load rating | 40 V DC, 100 mA Max
+
| Set-point External load rating || 40 V DC, 100 mA Max
 
|-
 
|-
| Gauge identification resistance | 33 kΩ ± 2%
+
| Gauge identification resistance || 33 kΩ ± 2%
 
|}
 
|}
  
Line 35: Line 33:
 
= References =
 
= References =
 
<ref> [https://www.idealvac.com/files/brochures/Edwards_APG_Active_Pirani_Gauges.pdf | Edwards_APG_Active_Pirani_Gauges.pdf, idealvac.com] </ref>
 
<ref> [https://www.idealvac.com/files/brochures/Edwards_APG_Active_Pirani_Gauges.pdf | Edwards_APG_Active_Pirani_Gauges.pdf, idealvac.com] </ref>
 +
 +
[[Category:Cross Refference]]

Revision as of 15:58, 20 December 2024

The APG-L is a early RJ45 connected active Pirani Gauge from BOC Edwards.

Specifications [1]
Maximum internal pressure: 6 bar absolute (5 bar gauge)
Pressure Range: 10 to 1 x 10⁻⁴ mbar
Volume of gauge tube: 6 cm³ (NW16/NW25) / 10 cm³ (15 mm OD)
Maximum operating altitude: 2000 m (Indoor use only)
Filament operating temperature (approximate): 100 ºC
Supply Voltage: +20 to +36V DC (<1V Ripple, <100 Ω Source)
Maximum power consumption: 1.5W
Signal Range: 2 ≤ output ≤ 10 V DC
Min. Load: 10kΩ
Set-point Adjustment Range: 2.2 to 8.9 V
Set-point External load rating 40 V DC, 100 mA Max
Gauge identification resistance 33 kΩ ± 2%

Voltage Pressure Table

References

[2]